Ono Sokki Linear Gauge Sensors (Long Stroke Type)
Code : GS-5050, GS-5100, GS-5051, GS-5101, GS-5050A, GS-5100A, GS-5051A, GS-5101A
Brand : Ono Sokki
Ono Sokki Linear Gauge Sensors (Long Stroke Type)
GS-5050, GS-5100, GS-5051, GS-5101, GS-5050A, GS-5100A, GS-5051A, GS-5101A
Long Stroke type
• Ball bearing is used.
• New models of resolution at 100 mm and resolution at 1 μm are added to line up.
• The pitch of mounting hole and the distance from tip of spindle after installation are same as those of the GS-5011.
|
Specifications
Model name
Item |
GS-5050 |
GS-5100 |
GS-5051 |
GS-5101 |
Measurement range |
50 mm |
100 mm |
50 mm |
100 mm |
Resolution |
10 μm |
1 μm |
Accuracy (at 20 °C) |
10 μm |
12 μm |
4 μm |
5 μm |
Maximum spindle velocity*1 |
1(4) m/s |
0.3(1.2) m/s |
Measurement force(downward)*2 |
3 N or less |
5.5 N less |
3 N or less |
5.5 N or less |
Number of sliding times
(proven in our endurance test) |
15 million times |
Protection class
(not including connector part) |
IP5X |
Stem diameter |
Ø15 mm+0-0.009 |
Power supply |
4.5 to 5.5 VDC |
Power consumption (5 VDC) |
120 mA or less |
Signal output (5 VDC) |
Two-phase square wave, phase difference: 90° ±20°
Output voltage Hi:4.5 V or more / Lo:0.4 V or less
|
Output impedance |
Approx. 22 Ω |
Vibration resistance
(when the power is off)*3 |
147 m/s2 in each of the three axial directions (75 minutes per each),
10 cycles of 10 Hz to150 Hz sweeping |
Shock resistance
(when the power is off)*3 |
1470 m/s2 in each of direction for each of three axes, three times for each of ±X,Y and Z directions, sine half-wave, pulse duration 6 ms |
Operating temperature range |
0 to +40 °C |
Storage temperature range |
-10 to +55 °C |
Cable length |
Approx. 4.9 m |
Weight (including cable, connector) |
Approx. 570 g |
Approx. 655 g |
Approx. 570 g |
Approx. 655 g |
Accessories |
Instruction manual, spanner |
*1. When used with Ono Sokki's Gauge Counter. The values within parentheses () are the maximum spindle velocity used with the DG-4140/DG-4190/DG-5100/DG-3000.
*2. An attachment may not return completely if it is facing upwards.
*3. Vibration/shock resistance values described in above are not guaranteed during measurement operation.
• Disconnected or modified signal cable is not applicable to CE marking.
|